Safety and accuracy in the clinical recording of cerebrospinal fluid pressure

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • Author(s): Nornes, H.; Magnæs, B.
  • Source:
    Acta Neurochirurgica; March 1973, Vol. 29 Issue: 1-2 p41-51, 11p
  • Additional Information
    • Abstract:
      Summary A pressure transducer with a disposable, presterilized fluid chamber was developed to increase the safety margin when recording cerebrospinal fluid (CSF) pressure. With this system the fluid compartment is completely separated from the transducer proper, reducing electrical hazard and the risk of infection. The operational range is form — 20 mm Hg to 300 mm Hg with good static accuracy. The transducer with stopcocks has a flat frequency response of up to 75 Hz, showing that the two-part principle is compatible with good dynamic performance.