Item request has been placed!
×
Item request cannot be made.
×
Processing Request
Particular issues to be considered in small field dosimetry for TrueBeam STx commissioning.
Item request has been placed!
×
Item request cannot be made.
×
Processing Request
- Additional Information
- Source:
Publisher: Pergamon Press Country of Publication: England NLM ID: 9306253 Publication Model: Print-Electronic Cited Medium: Internet ISSN: 1872-9800 (Electronic) Linking ISSN: 09698043 NLM ISO Abbreviation: Appl Radiat Isot Subsets: MEDLINE
- Publication Information:
Original Publication: Oxford ; New York : Pergamon Press, c1993-
- Subject Terms:
- Abstract:
This study aims to report the relevant issues concerning small fields in the commissioning of a TrueBeam STx for photon energies of 6MV, 10MV, 6FFF, and 10FFF. Percent depth doses, profiles, and field output factors were measured according to the beam model configuration of the treatment planning system. Multiple detectors were used based on the IAEA TRS-483 protocol as well as EBT3 radiochromic film. Analytical Anisotropic and Acuros XB algorithms, were configured and validated through basic dosimetry comparisons and end-to-end clinical tests.
Competing Interests: Declaration of competing interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
(Copyright © 2023 Elsevier Ltd. All rights reserved.)
- Contributed Indexing:
Keywords: Acuros algorithm; Field output factor; Small field dosimetry; TRS-483; TrueBeam STx
- Accession Number:
U68WG3173Y (Carmustine)
- Publication Date:
Date Created: 20231021 Date Completed: 20231103 Latest Revision: 20231103
- Publication Date:
20231103
- Accession Number:
10.1016/j.apradiso.2023.111066
- Accession Number:
37865066
No Comments.