Item request has been placed!
×
Item request cannot be made.
×
Processing Request
Current and Developing Technologies for BP Monitoring.
Item request has been placed!
×
Item request cannot be made.
×
Processing Request
- Author(s): Gunasekaran D;Gunasekaran D; Turner JM; Turner JM
- Source:
Current cardiology reports [Curr Cardiol Rep] 2023 Oct; Vol. 25 (10), pp. 1151-1156. Date of Electronic Publication: 2023 Sep 12.
- Publication Type:
Journal Article; Review
- Language:
English
- Additional Information
- Source:
Publisher: Current Science Country of Publication: United States NLM ID: 100888969 Publication Model: Print-Electronic Cited Medium: Internet ISSN: 1534-3170 (Electronic) Linking ISSN: 15233782 NLM ISO Abbreviation: Curr Cardiol Rep Subsets: MEDLINE
- Publication Information:
Original Publication: Philadelphia, PA : Current Science, c1999-
- Subject Terms:
- Abstract:
Purpose of Review: To discuss new and emerging technologies for blood pressure measurement and monitoring, including the limitations of current blood pressure measurement techniques, hopes for new device technologies, and the current barriers impeding change in this space.
Recent Findings: A number of new cuffless devices are being developed and poised to emerge on the marketplace in coming years. There are several different types of technologies and sensors currently under study. New guidelines for validation of cuffless blood pressure devices have recently been developed in anticipation of this change. The current standards for blood pressure device validation are specific to cuff-based technology and are insufficient for validating devices with cuffless-based technologies. In anticipation of a number of new cuffless technologies coming to market in the coming years, three sets of standards have been developed and published in recent years to address this gap.
(© 2023. The Author(s), under exclusive licence to Springer Science+Business Media, LLC, part of Springer Nature.)
- Contributed Indexing:
Keywords: Blood pressure validation standards; Cuffless blood pressure devices; Remote blood pressure monitoring