International Council for Standardization in Haematology (ICSH) recommendations for processing of blood samples for coagulation testing.

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • Additional Information
    • Source:
      Publisher: Blackwell Scientific Publications Country of Publication: England NLM ID: 101300213 Publication Model: Print-Electronic Cited Medium: Internet ISSN: 1751-553X (Electronic) Linking ISSN: 17515521 NLM ISO Abbreviation: Int J Lab Hematol Subsets: MEDLINE
    • Publication Information:
      Original Publication: Oxford : Blackwell Scientific Publications, c2007-
    • Subject Terms:
    • Abstract:
      This guidance document has been prepared on behalf of the International Council for Standardization in Haematology (ICSH). The aim of the document is to provide guidance and recommendations for the processing of citrated blood samples for coagulation tests in clinical laboratories in all regions of the world. The following areas are included in this document: Sample transport including use of pneumatic tubes systems; clots in citrated samples; centrifugation; primary tube storage and stability; interfering substances including haemolysis, icterus and lipaemia; secondary aliquots-transport, storage and processing; preanalytical variables for platelet function testing. The following areas are excluded from this document, but are included in an associated ICSH document addressing collection of samples for coagulation tests in clinical laboratories; ordering tests; sample collection tube and anticoagulant; preparation of the patient; sample collection device; venous stasis before sample collection; order of draw when different sample types are collected; sample labelling; blood-to-anticoagulant ratio (tube filling); influence of haematocrit. The recommendations are based on published data in peer-reviewed literature and expert opinion.
      (© 2021 John Wiley & Sons Ltd.)
    • References:
      Lippi G, Franchini M, Montagnana M, Salvagno GL, Poli G, Gudi GC. Quality and reliability of routine coagulation testing: can we trust that sample? Blood Coagul Fibrinolysis. 2006;17:513-519.
      Favaloro EJ, Lippi G, Adcock DM. Pre analytical and post analytical variables: the leading causes of diagnostic errors in haemostasis? Semin Thromb Hemost. 2008;34:612-634.
      Kalra J. Medical errors: impact on clinical laboratories and other clinical areas. Clin Biochem. 2004;37:1052-1062.
      Bonini P, Plebani M, Ceriotti F, Rubboli F. Errors in laboratory medicine. Clin Chem. 2002;48:691-698.
      Green SF. The cost of poor blood specimen quality and errors in preanalytical processes. Clin Biochem. 2013;46:1175-1179.
      Salvagno GL, Lippi G, Bassi A, Poli G, Guidi GC. Prevalence and type of pre analytical problems for in patient samples in coagulation laboratories. J Eval Clin Pract. 2008;14:351-353.
      Kitchen S, Adcock DM, Dauer R, et al. International Council for Standardisation in Haematology (ICSH) recommendations for collection of blood samples for coagulation testing. Int J Lab Hematol. 2021;43(4):571-580. 10.1111/ijlh.13584.
      Clinical Laboratory Standards Institute. Collection, Transport and Processing of Blood Specimens for Testing Plasma-Based Coagulation Assays And Molecular Haemostasis Assays: Approval Guidelines, 5th edition. CLSI H21-A510; 2008.
      Kim YA, Lewandrowski KB, Lucien FA, Van Cott EM. The effects of transport temperature and time on routine and specialized coagulation assays. Blood Coagul Fibrinolysis. 2018;29:184-188.
      Favaloro E, Nair SC, Forsyth CJ. Collection and transport of samples for laboratory testing in von Willbrand's disease (VWD): time for a reappraisal? Thromb Haem. 2001;86:1589-1590.
      Kitchen S, Malia RG, Greaves M, Preston FE. A method for the determination of activated factor VII using bovine and rabbit thromboplastins: demonstration of increased levels in disseminated intravascular coagulation. Thromb Res. 1988;50:191-200.
      Favaloro EJ, Soltani S, McDonald J. Potential laboratory misdiagnosis of hemophilia and von Willebrand disorder owing to cold activation of blood samples for testing. Am J Clin Pathol. 2004;122:686-692.
      Böhm M, Täschner S, Kretzschmar E, Gerlach R, Favaloro EJ, Scharrer I. Cold storage of citrated whole blood induces drastic time-dependent losses of factor VIII and von Willebrand factor: potential for misdiagnosis of haemophilia and von Willebrand disease. Blood Coagul Fibrinolysis. 2006;17:39-45.
      Refaii MA, van Cott EM, Lukoszyk M, Hughes J, Eby CS. Loss of factor VIII and von Willebrand activities during cold storage of whole blood is reversed by rewarming. Lab Hematol. 2006;12:99-102.
      Amukele TK, Sokoll LJ, Pepper D, Howard DP, Street J. Can unmanned aerial systems (Drones) be used for routine transport of chemistry, haematology and coagulation laboratory specimens? PLoS One. 2015;10(7):e0134020. 10.1371/jounal.pone.0134020.
      Kratz A, Salem RO, Van Cott EM. Effects of a pneumatic tube system on routine and novel haematology and coagulation parameters in healthy volunteers. Arch Pathol Lab Med. 2007;131:293-296.
      Dyszkiewicz-Korpanty A, Quinton R, Jassine J, Sarode R. The effect of pneumatic tube transport on PFA-100TM closure time and whole blood aggregation. J Thromb Haemost. 2004;2:354-356.
      Heireman L, Van Geel P, Musger L, Heylen E, Uyttenbroeck W, Mahieu B. Causes, consequences and management of sample haemolysis in the clinical laboratory. Clin Biochem. 2017;18:1317-1322.
      Gils C, Broell F, Vinholt PJ, Nielsn C, Nyobo M. Use of clinical data and acceleration profiles to validate pneumatic transportation systems. Clin Chem Lab med. 2020;58:560-568.
      Guss DA, Chan TC, Killeen JP. The impact of a pneumatic tube and computerised physician order management on laboratory turnaround time. Ann Emerg Med. 2008;51:181-185.
      Nybo M, Lund ME, Titlestad K, Maegaard CU. Blood sample transportation by pneumatic transportation systems: a systematic literature review. Clin Chem. 2018;64:782-790.
      Pragay DA, Edwards L, Toppin M, Palmer RR, Chilcote ME. Evaluation of an improved pneumatic tube system suitable for transportation of blood specimens. Clin Chem. 1974;20:57-60.
      Cui M, Jing R, Wang H. Changes of serum lactate dehydrogenase and potassium levels produced by a pneumatic tube system. Lab Med. 2009;40:728-731.
      Gomez-Rioja R, Fernandez-Calle P, Alcaide MJ, et al. Inter-individual variability of hemolysis in plasma samples during pneumatic tube system transport. Clin Chem Lab Med. 2013;51:e231-e233.
      Helms CC, Marvel M, Zhao W, et al. Mechanisms of hemolysis-associated platelet activation. J Thromb Haemost. 2013;11:2148-2154.
      Streichert T, Otto B, Schnabel C, et al. Determination of hemolysis thresholds by the use of data loggers in pneumatic tube systems. Clin Chem. 2011;57:1390-1397.
      ISO 15189. Medical laboratories - Requirements for quality and competence; 2012.
      Farnsworth CW, Webber DM, Krekler JA, Budelier MM, Bartlett NL, Gronowski AM. Parameters for validating a hospital pneumatic tube system. Clin Chem. 2019;65:694-702.
      Cooper J, der Van Hoop J, Broell F, Kitchen S. Identifying and reducing high levels of haemolysis from samples transport in pneumatic tube systems. J Thromb Haemost. 2020. In press.
      Delbruck C, Miesbach W. The course of von Willebrand factor and factor VIII activity in patients with von Willebrand disease during pregnancy. Acta Haematol. 2019;142:71-78.
      Austin AW, Wirtz PH, Patterson SM, Stutz M, von Kanel R. Stress-induced alterations in coagulation: assessment of a new hemoconcentration correction technique. Psychosom Med. 2012;74:288-295.
      Bowyer A, Kitchen S, Maclean R. Effects of emicizumab on APTT, one-stage and chromogenic assays of factor VIII in artificially spiked plasma and in samples from haemophilia A patients with inhibitors. Haemophilia. 2020;26:536-542.
      Lippi G, Salvagno GL, Monyagna M, Manzaato F, Guidi GC. Influence of the centrifuge time of primary plasma tubes on routine coagulation testing. Blood Coag Fibrinol. 2007;18:525-528.
      Carroll WE, Wollitzer AO, Harris L, Ling MC, Whitaker WL, Jackson RD. The significance of platelet count in coagulation studies. J Med. 2001;32:83-86.
      Nelson S, Pratt A, Marlar RA. Rapid preparation of plasma for ‘stat’ coagulation testing. Arch Pathol Lab Med. 1994;118:175-176.
      Kao CH, Shy LC, Yen WF. Evaluation of a high-speed centrifugation with rapid preparation of plasma for coagulation testing to improve turnaround time. J Biomed Lab Sci. 2010;22:23-27.
      Pengo V, Tripodi A, Reber G, et al. Subcommittee on lupus anticoagulant/antiphospholipid antibody of the scientific and standardisation committee of the International Society on Thrombosis and Haemostasis. Update of the guidelines for lupus anticoagulant detection. J Thromb Haemost. 2009;7:1737-1740.
      Tripod A, Valsecchi C, Chantarangkul V, Battaglioli MPM. Standardisation of activated protein C resistance testing: effects of residual platelets in frozen plasmas assessed by commercial and home-made methods. Br J Haem. 2003;120:825-828.
      Favaloro EJ. Pre analyticial variables in coagulation testing. Blood Coagul Fibrinolysis. 2007;18:86-89.
      Baglin T, Luddington R. Reliability of delayed INR determination: implications for decentralised anticoagulant care with off-site blood sampling. Br J Haem. 1997;96:431-434.
      Heil W, Grunewals R, Amnd M, Heins M. Influence of time and temperature on coagulation analysis in stored plasma. Clin Chem Lab Med. 1998;36:459-452.
      Awad MA, Selim TE, Al-Sabbagh FA. Influence of storage time and temperature on International Normalized Ratio (INR) levels and plasma activities of vitamin K dependent clotting factors. Hematology. 2004;9:333-337.
      O'Neill EM, Rowley J, Hanson-Wicher H, McCarter S, Ragno G, Valeri CR. Effect of 24-hour whole-blood storage on plasma clotting factors. Transfusion. 1999;39:488-491.
      Rao LV, Okorodudu AO, Petersen JR, Elghetany MT. Stability of prothrombin time and activated partial thromboplastin time tests under different storage conditions. Clin Chim Acta. 2000;300:13-21.
      van Geest-Daalderop JH, Mulder AB, Boonman-deWinter LJ, Hoekstra MM, van den Besselaar AM. Preanalytical variables and off-site blood collection: influences on the results of the prothrombin time/international normalized ratio test and implications for monitoring oral anticoagulant therapy. Clin Chem. 2005;51:561-568.
      Toulon P, Metge S, Hangard M, Zwahlen S, Paiulenne S, Besson V. Impact of different storage times at room temperature of unspun citrated blood samples on routine coagulation test results. Results of a bicentre study and review of the literature. Int J Lab Hematol. 2017;39:458-468.
      Zurcher M, Sulzer I, Barizzi G, Lammle B, Alberio L. Stability of coagulation assays performed in plasma from citrated whole blood transported at ambient temperature. Thromb Haemost. 2008;99:416-426.
      Linskens EA, Devreese KMJ. Pre-analytical stability of coagulation parameters in plasma stored at room temperature. Int J Lab Hematol. 2018;40:292-303.
      Caliezi C, Reber G, Lämmle B, de Moerloose P, Wuillemin WA. Agreement of D-dimer results measured by a rapid ELISA (VIDAS) before and after storage during 24 hr or transportation of the original whole blood samples. Thromb Haemost. 2000;83:177-178.
      Adcock DA, Kressin DC, Marlar RA. The effect of time and temperature variables on routine coagulation tests. Blood Coagul Fibrinolysis. 1998;9:463-470.
      Birri N, Baumgartner D, Conte T, et al. Stability of low molecular weight heparin anti-factor Xa activity in citrated whole blood and plasma. Br J Haematol. 2011;155:620-641.
      Adcock Funk DM, Lippi G, Favaloro EJ. Quality Standards for sample processing transportation and storage in haemostasis testing. Semin Thromb Hemost. 2012;38:576-585.
      Rimac V, Coen HD. Is it acceptable to use coagulation plasma samples stored at room temperature and 4oC for 24 hours for additional PT APTT fibrinogen, antithrombin and D-dimer testing? Int J Lab Hematol. 2017;39:475-481.
      Lippi G, Guidi GC, Mattiuzzi C, Plebani M. Pre-analytical variability: the dark side of the moon in laboratory testing. Clin Chem Lab Med. 2006;44:358-365.
      Tantanate C, Teyateeti M, Tiendtadakul P. Influence of plasma interferences on screening coagulation and performance evaluation of the automated coagulation analyser Sysmex CS2100. Siriraj Med J. 2010;63:151-156.
      Lippi G, Ippolito L, Favaloro EJ. Technical evaluation of the novel pre-analytical module on instrumentation laboratory ACL TOP: advancing automation in hemostasis testing. J Lab Autom. 2013;18:382-390.
      Florin L, Oyaert M, Van Maerken T, Devreese KMJ. Performance evaluation of the pre-analytical check module of the Stago STA R Max2 mechanical endpoint detection analyzer for the assessing of the impact of hemolysis lipemia and icterus on APTT and PT. Int J Lab Hematol. 2018;40:e109-e112.
      Montanuli B, Guitto C, Cosseddu D. Influence of hemolysis, icterus and lipemia on coagulation tests as performed on Cobas t511 new analyser. Blood Coag Fibrinol. 2020;31:48-54.
      Woolley A, Golmard JL, Kitchen S. Effects of haemolysis, icterus and lipaemia on coagulation tests as performed on Stago STA-Compact-Max analyser. Int J Lab Hematol. 2016;38:375-388.
      Duhalde H, Skogo J, Karlsson M. Point-of-Care hemolysis detection in blood gas specimens directly at the emergency department. Scan J Clin Lab Investig. 2019;79:283-287.
      Lippi G, Montagnana M, Salvagno GL, Guidi GC. Interference of blood cell lysis on routine coagulation testing. Arch Pathol Lab Med. 2006;130:181-184.
      Lippi G, Avanzini P, Zobbi V, Ippolito L. Influence of mechanical hemolysis of blood on two D-dimer immunoassays. Blood Coagul Fibrinolysis. 2012;23:461-463.
      Lippi G, Plebani M, Favaloro EJ. Interference in coagulation testing: focus on spurious hemolysis, icterus, and lipemia. Semin Thromb Hemost. 2013;39:258-266.
      Laga AC, Cheves TA, Sweeney JD. The effect of specimen hemolysis on coagulation test results. Am J Clin Pathol. 2006;126:748-754.
      Woolley A, Horner K, Kitchen S. Haemolysis during sample collection causes variable but sometimes clinically relevant important errors. J Thromb Haemost. 2011;9(Suppl. 2):125.
      Mackie I, Mancini I, Muia J, et al. International Council for Standardization in Hematology (ICSH) recommendations for laboratory measurement of ADAMTS13. Int J Lab Hematol. 2020;42:685-696.
      Nagant C, Rozen L, Demulder A. HIL interferences on three hemostasis analyzers and contribution of a preanalytical module for routine coagulation assays. Clin Lab. 2016;62:1979-1987.
      Dimeski G, Jones BW. Lipemic samples: effective process for lipid reduction using high speed centrifugation compared with ultracentrifugation. Biochem Med. 2011;21:86-94.
      Negrini D, Bernadi D, Antonelli G, Plebani M. Interference of lipemia in samples for routine coagulation testing using a Sysmex CS-5100 coagulometer. Int J Lab Hematol. 2019;41:772-777.
      Gardiner C, Lane P, Tailor H, Mackie IJ. A practical method for reducing the interference due to lipemia in coagulation tests. Int J Lab Hematol. 2020;42:140-144.
      van Rossum AP, Vlasveld LT, van den Hoven LJ, de Wit AWM, Castel A. False prolongation of the APTT in inflammatory patients: interference of C-reactive protein. Br J Haematol. 2012;57:381-440.
      Schouwers SM, Delanghe JR, Devreese KM. Lupus Anticoagulant (LAC) testing in patients with inflammatory status: does C-reactive protein interfere with LAC test results? Thromb Res. 2010;125:102-104.
      Lippi G, Salvagno GL, Adcock DM, Gelati M, Guidi GC, Favaloro EJ. Right or wrong sample received for coagulation testing? Tentative algorithms for detection of an incorrect type of sample. Int J Lab Hematol. 2010;32:132-137.
      Gosselin RC, Honeychurch K, Kang HJ, Dwyre DM. Effect of multiple freeze-thaw cycles on coagulation testing. Semin Thromb Hemost. 2020;46:515-520.
      Woodhams B, Giradot O, Blanco M, Collesse G, Gourmelin Y. Stability of coagulation proteins in frozen plasma. Blood Coagul Fibrinolysis. 2001;12:229-236.
      Gosselin RC, Dwyre DW. Determining the effect of freezing on coagulation testing: comparison of results between fresh and once frozen-thawed plasma. Blood Coagul Fibrinolysis. 2015;26:69-74.
      Vis R, van den Besselaar AMHP, Witteveen E, Meeuwise-Braun J, Schafer-van MH. Artificial prolongation of the prothrombin time of lyophilised plasma induced by transportation with solid carbon dioxide. Thromb Haemost. 1992;1:725.
      Trondsetås L, Mikkelsen G, Lian IA. The effects of dry ice exposure on plasma pH and coagulation analyses. Clin Chem Lab Med. 2017;56:59-64.
      Odsaeter IH, Lian IA, Bratberg K, Mikkelsen G. Dry ice exposure of plasma samples influences pH and lupus anticoagulant analysis. Chem Lab Med. 2015;53:809-813.
      Jo S-Y, Kim I-S, Kim H. Potential misdiagnosis of von Willebrand disease caused by insufficient mixing after thawing of froizen plasma. Int J Lab Hematol. 2020;43:831-836.
      Zhao Y, Feng G, Feng L. Effects of pre analytical storage time, temperature and freeze-thaw time son coagulation factor activities in citrate-anticoagulated plasma. Ann Transl Med. 2018;6:456-465.
      Maelegheer K, Devreese KMJ. The impact of repeated freeze-thaw cycles on antiphospholipd antibody titer. Res Pract Thromb Haemost. 2018;6:366-369.
      Clinical and Laboratory Standards Institute. Platelet Function Testing by Aggregometry: Approved Guideline. Clinical and Laboratory Standards Institute; 2008;28:1-45.
      Harrison P, Mackie I, Mumford A, et al. Guidelines for the laboratory investigation of heritable disorders of platelet function. Br J Haematol. 2011;155:30-44.
      Cattaneo M, Cerletti C, Harrison P, et al. Recommendations for the Standardization of Light Transmission Aggregometry: A Consensus of the Working Party from the Platelet Physiology Subcommittee of SSC/ISTH. J Thromb Haemost. 2013;11:1183-1189.
      Le Quellec S, Bordet J-C, Negrier C, Dargaud Y. Comparison of current platelet functional tests for the assessment of aspirin and clopidogrel response. A review of the literature. Thromb Haemost. 2016;116:638-650.
      Kottke-Marchant K, Corcoran G. The laboratory diagnosis of platelet disorders. Arch Pathol Lab Med. 2002;126:133-146.
      Germanovich K, Femia EA, Cheng CY, Dovlatova N, Marco CM. Effects of pH and concentration of sodium citrate anticoagulant on platelet aggregation measured by light transmission aggregometry induced by adenosine diphosphate. Platelets. 2018;29:21-26.
      Stegnar M, Knezevic A, Bozic-Mijovski M. The effect of pre-analytical variables on light transmittance aggregometry in citrated platelet-rich plasma from healthy subjects. Clin Chem Lab Med. 2010;48:1463-1465.
      Andreasen JB, Pistor-Riebold TU, Knudsen IH, Ravn HB, Anne-Hvas AM. Evaluation of different sized blood sampling tubes for thromboelastometry, platelet function, and platelet count. Clin Chem Lab Med. 2014;52:701-706.
      Hellstern P, Sturzebecher U, Wuchold B, et al. Preservation of in vitro function of platelets stored in the presence of a synthetic dual inhibitor of factor Xa and thrombin. J Thromb Haemost. 2007;5:2119-2126.
      Haubelt H, Vogt A, Hellstern P. Preservation of platelet aggregation and dense granule secretion during extended storage of blood samples in the presence of a synthetic dual inhibitor of factor Xa and thrombin. Platelets. 2008;19:496-501.
      Tauer JT, Lohse J, Konig S, Knofler R. Investigations of platelet function in whole blood with BAPA as anticoagulant. Transfus Med Hemother. 2010;37:284-288.
      Mani H, Hellis M, Lindhoff-Last E. Platelet function testing in hirudin and BAPA anticoagulated blood. Clin Chem Lab Med. 2011;49:501-507.
      Chapman K, Favaloro EJ. Time dependent reduction in platelet aggregation using the multiplate analyser and hirudin blood due to platelet clumping. Platelets. 2018;29:305-308.
      Favaloro EJ. Utility of the PFA-100 for assessing bleeding disorders and monitoring therapy: a review of analytical variables, benefits and limitations. Haemophilia. 2001;7:170-179.
      Jilma B. Platelet function analyser (PFA-100): a tool to quantify congenital or acquired platelet dysfunction. J Lab Clin Med. 2001;138:152-163.
      Heilmann EJ, Kundu SK, Sio R, Garcia C, Gomez R, Christie DJ. Comparison of four commercial citrate blood collection systems for platelet function analysis by the PFA-100 system. Thromb Res. 1997;87:159-164.
      Harrison P, Robinson MSC, Mackie IJ, et al. Performance of the platelet function analyser PFA-100R in testing abnormalities of primary haemostasis. Blood Coagul Fibrinolysis. 1999;10:25-31.
      Welch EL, Crooks MG, Hart SP. Agreement between blood draw techniques for assessing platelet activation by flow cytometry. Platelets. 2019;30:530-534.
      Schmitz G, Rothe G, Ruf A, et al. European working group on clinical cell analysis: consensus protocol for the flow cytometric characterisation of platelet function. Thromb Haemost. 1998;79:885-896.
      Goodall AH, Appleby J. Flow-cytometric analysis of platelet membrane glycoprotein expression and platelet activation. Methods Mol Biol. 2004;272:225-253.
      Michelson AD, Linden M, Barnard MR, Furman MI, Frelinger AL. Flow cytometry. In: Michelson AD, ed. Platelets. Academic Press; 2007: 545-563.
      Lance MD, Henskens YMC, Nelemans P, et al. Do blood collection methods influence whole-blood platelet function analysis? Platelets. 2013;24:275-281.
      Le Quellec S, Paris M, Nougier C, et al. Pre-analytical effects of pneumatic tube system transport on routine haematology and coagulation tests, global coagulation assays and platelet function assays. Thromb Res. 2017;153:7-13.
      Poletaev AV, Koltsova EM, Ignatova AA, et al. Alterations in the parameters of classic, global, and innovative assays of hemostasis caused by sample transportation via pneumatic tube system. Thromb Res. 2018;170:156-164.
      Wallin O, Soderberg J, Grankvist K, Andreas Jonsson P, Hultdin J. Pre-analytical effects of pneumatic tube transport on routine haematology, coagulation parameters, platelet function and global coagulation. Clin Chem Lab Med. 2008;46:1443-1449.
      Thalen S, Forsling I, Eintrei J, Soderblom L, Antovic JP. Pneumatic tube transport affects platelet function measured by multiplate electrode aggregometry. Thromb Res. 2013;132:77-80.
      Femia EA, Pugliano M, Podda G, Cattaneo M. Comparison of different procedures to prepare platelet-rich plasma for studies of platelet aggregation by light transmission aggregometry. Platelets. 2012;23:7-10.
      Merolla M, Nardi MA, Berger JS. Centrifugation speed affects light transmission aggregometry. Int J Lab Hematol. 2012;34:81-85.
      Cattaneo M, Lecchi A, Zighetti ML, Lussana F. Platelet aggregation studies: autologous platelet-poor plasma inhibits platelet aggregation when added to platelet-rich plasma to normalize platelet count. Haematologica. 2007;92:694-697.
      Cadamuro J, Simundic AM, Ajner E, Sandberg S. A pragmatic approach to sample acceptance and rejection. Clin Biochem. 2017;50:579-581.
    • Contributed Indexing:
      Keywords: ICSH; coagulation; sample processing
    • Publication Date:
      Date Created: 20210928 Date Completed: 20220203 Latest Revision: 20220203
    • Publication Date:
      20231215
    • Accession Number:
      10.1111/ijlh.13702
    • Accession Number:
      34581008