The traceability of commercial plasma calibrators to the plasma International Standards for factor VIII and factor IX.

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • Author(s): Wilmot HV;Wilmot HV; Rakowski K; Rakowski K; Gray E; Gray E
  • Source:
    International journal of laboratory hematology [Int J Lab Hematol] 2020 Dec; Vol. 42 (6), pp. 810-818. Date of Electronic Publication: 2020 Jul 08.
  • Publication Type:
    Journal Article
  • Language:
    English
  • Additional Information
    • Source:
      Publisher: Blackwell Scientific Publications Country of Publication: England NLM ID: 101300213 Publication Model: Print-Electronic Cited Medium: Internet ISSN: 1751-553X (Electronic) Linking ISSN: 17515521 NLM ISO Abbreviation: Int J Lab Hematol Subsets: MEDLINE
    • Publication Information:
      Original Publication: Oxford : Blackwell Scientific Publications, c2007-
    • Subject Terms:
    • Abstract:
      Accurate measurement of coagulation factors is essential, especially for diagnosis of deficiency. Clinical laboratories use commercially available plasma calibrators, which should be traceable to the relevant plasma International Standard (IS). This study assessed the relationship between the plasma IS for factors IX (FIX) and VIII (FVIII) and some commonly used commercial calibrators. Calibrators from seven manufacturers were assayed for FIX and FVIII activity by one-stage clotting assay (OSCA) using different activated partial thromboplastin time (APTT) reagents and deficient plasmas, or chromogenic assay (CA). Results were calculated relative to the 4th IS Factors II,VII,IX,X, Plasma or the 6th IS Factor VIII/VWF, Plasma. Results for each calibrator were similar across the APTT reagents and deficient plasmas used. All calibrators showed a recovery of 90%-111% of the manufacturers' values, except calibrator C, which had recovery of around 85%. CA gave similar results, with good recovery for all but calibrator C. Similar low recoveries for OSCA and CA were found for a different lot of calibrator C and for a different calibrator product from manufacturer C. When all calibrators from manufacturer C were assayed by OSCA using the manufacturer's own deficient plasmas and APTT reagents, the mean recovery was still below 90%. Overall, there was good traceability of the international unit between the IS and commercial calibrator plasmas. Calibrators from one manufacturer consistently yielded lower than expected values for FIX and FVIII. This could lead to an over-estimation of the coagulation factor content in patient samples and demonstrates the importance of careful choice of calibrator.
      (© 2020 The Authors. International Journal of Laboratory Hematology published by John Wiley & Sons Ltd.)
    • References:
      J Thromb Haemost. 2019 Apr;17(4):567-573. (PMID: 30667575)
      Transfus Apher Sci. 2018 Aug;57(4):480-484. (PMID: 30093248)
      Thromb Haemost. 2001 Mar;85(3):560. (PMID: 11307831)
      Eur J Haematol. 2015 Feb;94 Suppl 77:38-44. (PMID: 25560793)
      Semin Thromb Hemost. 2007 Apr;33(3):283-9. (PMID: 17427063)
      Blood Coagul Fibrinolysis. 2011 Mar;22(2):151-4. (PMID: 21192255)
      Haemophilia. 2015 Jul;21(4):550-7. (PMID: 25688866)
      Haemophilia. 2019 Jan;25(1):162-169. (PMID: 30488994)
      J Thromb Haemost. 2011 Oct;9(10):2100-2. (PMID: 21848692)
      Semin Thromb Hemost. 2007 Apr;33(3):250-8. (PMID: 17427059)
      Int J Lab Hematol. 2020 Dec;42(6):810-818. (PMID: 32638532)
      Haemophilia. 2018 Sep;24(5):e363-e368. (PMID: 30051554)
    • Contributed Indexing:
      Keywords: Calibrator discrepancy; FIX/FVIII activity assays; International Standards; International Unit traceability; Plasma calibrators
    • Accession Number:
      839MOZ74GK (F8 protein, human)
      9001-27-8 (Factor VIII)
      9001-28-9 (Factor IX)
    • Publication Date:
      Date Created: 20200709 Date Completed: 20210203 Latest Revision: 20210203
    • Publication Date:
      20240829
    • Accession Number:
      PMC7754292
    • Accession Number:
      10.1111/ijlh.13277
    • Accession Number:
      32638532