Clinical Application of Custom Neck Collar with Negative Sensory Feedback in Children with Intractable Torticollis.
Item request has been placed!
×
Item request cannot be made.
×
Processing Request
- Additional Information
- Subject Terms:
- Abstract:
Background/Objectives: The aim of this study was to investigate the effect of a custom neck collar with negative sensory feedback for the treatment of torticollis that was previously unresponsive to conservative or surgical treatment. Methods: Twenty-four children diagnosed with unresponsive or intractable torticollis were enrolled in this two-stage, single-arm study. The ipsilateral aspect of the orthosis is adjustable in height and designed to provide support between the clavicle and the mandibular angle on the tilted side. In stage 1 (the adjustment period), the orthosis with a smooth surface was applied for 2 h per day for 3 months. In stage 2, a rough surface with a hook-and-loop fastener (Velcro©) was attached to the collar, and it was worn for a further 2 h a day for 3 months. Twenty children (mean age 63.95 ± 13.44 months) were included in the analysis. Results: The mean torticollis angle was 17.60 ± 5.65° (mean ± SD) at baseline; 14.15 ± 3.62° directly after stage 1; and 6.00 ± 3.67° directly after stage 2 (X2 = 36.685, df = 19, p = 0.000). Conclusions: This study demonstrated the feasibility, therapeutic effect, and safety of a novel tactile feedback orthosis for the treatment of children with torticollis. The use of a custom neck collar with negative sensory feedback may be a viable therapeutic option for the treatment of unresolved or intractable torticollis. [ABSTRACT FROM AUTHOR]
- Abstract:
Copyright of Children is the property of MDPI and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
No Comments.