Possibilities of the Determination of Matrix Elements of High-Purity Glasses of As-Se and As-S Systems by Inductively Coupled Plasma Atomic Emission Spectrometry.

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • Additional Information
    • Abstract:
      Possibilities of inductively coupled plasma atomic emission spectrometry (ICP AES) as an alternative to X-ray spectrochemical analysis are estimated in the determination of matrix elements of glasses of As-Se and As-S systems. Standard solutions were prepared from pure elements As, Se and S. It was shown that the matrix elements of glasses can be determined with an expanded uncertainty of 0.05-0.1 mol %. The results of determinations by ICP AES and X-ray fluorescence spectrometry were compared; and it was found that by the attained performance characteristics, the results of ICP AES are highly competitive with the data of X-ray fluorescence spectrometry, but do not require sets of adequate solid reference samples. [ABSTRACT FROM AUTHOR]
    • Abstract:
      Copyright of Journal of Analytical Chemistry is the property of Springer Nature and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)